A comparative surface evaluation of orthodontic mini-implants before and after en masse retraction—A SEM study

Bibliographic Details
Title: A comparative surface evaluation of orthodontic mini-implants before and after en masse retraction—A SEM study
Authors: Alok Ranjan, Pravin Shetty, Rahul Despande, Anil Biradar, Waseem Khan, Rohit Kulshrestha
Source: Journal of Orthodontic Science, Vol 12, Iss 1, Pp 15-15 (2023)
Publisher Information: Wolters Kluwer Medknow Publications, 2023.
Publication Year: 2023
Collection: LCC:Dentistry
Subject Terms: en-masse retraction, micro-implants, scanning electron microscopy, Dentistry, RK1-715
More Details: OBJECTIVE: To evaluate the changes in surface morphology of two different types of mini-implants after clinical en masse retraction using scanning electron microscopy. MATERIAL AND METHODS: Fifty mini-implants of Dentos (Korea, Absoanchor, BH-1817-08) and Orlus (Korea, Yesanchor, C-1817) were inserted in patients in a split-mouth design who required en masse anterior retraction and absolute anchorage. Surface characteristics of mini-implants such as pitch (distance between consecutive threads), flank width (distance between root and crest), and taper were studied using scanning electron microscope (FEI nanosem450) before and after clinical use. RESULTS: Statistically significant difference (p value = 0.003) was found in a mean reduction of pitch dimension among the two groups with a mean difference of 25.000 μm. Also, a statistically significant difference was noted (p value = 0.001) in a mean reduction of flank width among Dentos implants as compared to Orlus implants. A statistically significant difference (p = 0.001) was seen in the mean reduction of taper dimension among Dentos group (0.0140 ± 0.02271) as compared to the Orlus group (0.0810 ± 0.05152). CONCLUSION: A marked reduction in surface morphology such as the pitch, flank width, and taper of both mini-implants after retrieval was observed. Dentos group of mini-implants displayed better dimensional stability post-retrieval as compared to the Orlus group of mini-implants. All the mini-screws showed milling defects in form of scratches on observation under scanning electron microscopy despite a smooth appearance to the naked eye.
Document Type: article
File Description: electronic resource
Language: English
ISSN: 2278-0203
Relation: http://www.jorthodsci.org/article.asp?issn=2278-0203;year=2023;volume=12;issue=1;spage=15;epage=15;aulast=Ranjan; https://doaj.org/toc/2278-0203
DOI: 10.4103/jos.jos_166_21
Access URL: https://doaj.org/article/87d55b6daaae4f50b7133502cdede6e5
Accession Number: edsdoj.87d55b6daaae4f50b7133502cdede6e5
Database: Directory of Open Access Journals
More Details
ISSN:22780203
DOI:10.4103/jos.jos_166_21
Published in:Journal of Orthodontic Science
Language:English