Impact of using cross-over CV and mean for two different lots of assay control on implementation of Westgard rules in chemical diagnostic tests

Bibliographic Details
Title: Impact of using cross-over CV and mean for two different lots of assay control on implementation of Westgard rules in chemical diagnostic tests
Authors: Ayman Mohamed Nabil, Hayat Mirza Alsaif, Muneer Ahmad Aljamaan, Abdullah Abdullah H. Algafly, Rashad Hassan aleid, Raji Ali Helal, Zainab Ali Hussain almutawah, Amani Abdulkareem S. Alzayer, Walaa Ali Hussain Almutawah, Badr Abdullah Motlaq AlKhalaf
Source: Practical Laboratory Medicine, Vol 44, Iss , Pp e00449- (2025)
Publisher Information: Elsevier, 2025.
Publication Year: 2025
Collection: LCC:Medicine (General)
LCC:Chemistry
Subject Terms: Quality control, Westgard rule, Cross-over coefficient, Accuracy, Diagnostic test reliability, Reagent lot variation, Medicine (General), R5-920, Chemistry, QD1-999
More Details: Background: The main challenges of clinical laboratories concerning quality control include cost-effectiveness, variability in standardized materials, and evolving technologies across various diagnostic fields. While traditional QC practices and automation systems provide for accuracy, gaps exist, especially when applying Westgard rules to control lots for multiple assays. Such gaps result in inconsistent QC outcomes and unaddressed challenges in diagnostic reliability. Objective: This study aims to assess the effect of the cross-over coefficient of variation (CV) and mean values for different assay control lots on implementing Westgard rules to improve QC practices and enhance the accuracy and reliability of diagnostic tests in molecular laboratories. Methods: Data from 18 Levy-Jennings charts, with two assay control lots, were analyzed. Statistical comparisons of failure rates before and after setting the actual SD were performed using chi-square or T-tests at p
Document Type: article
File Description: electronic resource
Language: English
ISSN: 2352-5517
Relation: http://www.sciencedirect.com/science/article/pii/S2352551725000022; https://doaj.org/toc/2352-5517
DOI: 10.1016/j.plabm.2025.e00449
Access URL: https://doaj.org/article/a0d8b88f17bd46dfab385fc2bd3cee61
Accession Number: edsdoj.0d8b88f17bd46dfab385fc2bd3cee61
Database: Directory of Open Access Journals
More Details
ISSN:23525517
DOI:10.1016/j.plabm.2025.e00449
Published in:Practical Laboratory Medicine
Language:English