Automated Proton Spectroscopic Image Processing

Bibliographic Details
Title: Automated Proton Spectroscopic Image Processing
Authors: Doyle, T.J., Pathak, R., Wolinsky, J.S., Narayana, P.A.
Source: Journal of Magnetic Resonance - Series B; January 1995, Vol. 106 Issue: 1 p58-63, 6p
Abstract: An automatic data processing and display protocol for proton magnetic resonance spectroscopic imaging is described. The spectroscopic images are generated using the fitted peak areas of phased metabolite resonances using the Levenberg-Marquardt algorithm. This automated spectroscopic image processing is tested and evaluated using phantom and human brain data. This algorithm is robust, simple to use, and minimizes user bias in processing and quantifying serial changes of metabolite levels in brain tissue.Copyright 1995, 1999 Academic Press,Inc.
Database: Supplemental Index
More Details
ISSN:10641866
10960872
DOI:10.1006/jmrb.1995.1008
Published in:Journal of Magnetic Resonance - Series B
Language:English