Normal Gözlerde Goldmann Applanasyon Tonometre ile Rebound Tonometri, Dinamik Kontur Tonometre, Tonopen XL'ın Karşılaştırılması.
Title: | Normal Gözlerde Goldmann Applanasyon Tonometre ile Rebound Tonometri, Dinamik Kontur Tonometre, Tonopen XL'ın Karşılaştırılması. |
---|---|
Alternate Title: | Comparison of Rebound Tonometer, Dynamic Contour Tonometer, Tonopen XL with Goldmann Applanation Tonometer in Normal Eyes. |
Authors: | ÖZCAN, Altan Atakan1 altanoz@gmail.com, ÇİLOĞLU, Emine2 drciloglu@gmail.com, ESEN, Ebru3 ebrublg@yahoo.com, ÜNDAR, İsmail Hakkı4 dr_ismailundar@hotmail.com, HARBİYELİ, İnan5 inanharbiyeli@hotmail.com, ŞİMŞEK, Firas6 firassimsek@gmail.com |
Source: | Journal of Glaucoma-Cataract / Glokom-Katarakt. Dec2013, Vol. 8 Issue 4, p225-230. 6p. 3 Charts. |
Subject Terms: | *MEDICAL equipment, *TONOMETERS, *COMPARATIVE studies, *THICKNESS measurement, *ULTRASONICS, *STANDARD deviations |
Abstract (English): | Purpose: This clinical study was conducted to compare intraocular pressure (IOP) measurements by Goldmann applanation tonometry (GAT), Rebound tonometry (Icare), Dynamic Contour tonometer (Pascal) and TonoPen XL (TP), to evaluate the effect of central corneal thickness (CCK) on IOP measurements. Materials and Methods: The study included 39 healthy subjects. IOP measurements were taken using Icare, Pascal, TP and GAT, respectively. CCK was measured with ultrasound pachymetry. To compare the measurements; the analysis of percentage change and Pearson correlation, the average, median, standart deviation and minimum-maximum values were calculated. The agreement between 2 different devices was assessed by use of the Bland-Altman method. Results: The average CCK was 528±16 μm (483-555 μm). The mean IOP, with Icare was 13.8±2.1 mmHg, with Pascal DCT was 14.0±2.2 mmHg, with TP was 14.0±3.4 mmHg, with GAT was 13.8±2.4 mmHg. There was no significant difference statistically between the IOP measurements of all methods (p>0.05). The percentage of the Icare, Pascal DCT and TP readings, which were all within ±3 mmHg of the GAT readings were respectively 66.6%, 61.5%, 56.4%. There was an agreement between corrected GAT values and the other tonometry values on the Bland-Altman graphs. There was no correlation between IOP measurements and CCT. Conclusion: In our study, there was no significant difference between the four methods of IOP measurements. Icare tonometry does not require topical anesthesia, is easy to use, the measurement is fast. So, it can be used as a relevant tonometry in clinical setting. [ABSTRACT FROM AUTHOR] |
Abstract (Turkish): | Amaç: Bu klinik çalışmada Goldmann applanasyon tonometresi (GAT), Rebound tonometre (İcare), Dinamik kontur tonometre (Pascal DKT) ve TonoPen XL (TP) ile yapılan göz içi basınç ölçümlerinin (GİB) karşılaştırılması ve İcare, Pascal DKT, TP ile yapılan göz içi basınç ölçümleri üzerine santral kornea kalınlığının etkisini değerlendirmek amacıyla planlanmıştır. Gereç ve Yöntem: Otuz dokuz sağlıklı birey çalışmaya dahil edildi. Olguların sırasıyla İcare, Pascal DKT, TP ve GAT ile GİB ölçümleri yapıldı. Ultrasonik pakimetri kullanılarak santral kornea kalınlığı (SKK) değerleri elde edildi. Ölçümlerin karşılaştırılması için yüzde değişim analizi, Pearson korelasyon analizi ile ortalama, standart sapma değerleri ve minimum-maksimum değerleri hesaplandı. İki farklı metod arasındaki uyumluluk Bland-Altman metodu kullanılarak değerlendirildi. Bulgular: Çalışmaya dahil edilen olguların ortalama yaşı 38±4 (30-45) idi. Ortalama SKK 528±16 μm (483-555 μm). Ortalama GİB; İcare ile ortalama 13.8±2.1 mmHg, Pascal DKT ile ortalama 14.0±2.2 mmHg,TP ile 14.0±3.4 mmHg, GAT ile ortalama 13.8±2.4 mmHg, olarak ölçüldü. Ölçümler arasında anlamlı farklılık görülmedi. İcare, Pascal DKT, TP ile yapılan ölçümlerin yüzde kaçının GAT ile elde edilen değerlerin ±3 mmHg aralığında olduğu araştırıldığında sırasıyla %66.6, %61.5, %56.4 değerler elde edildi. Bland-Altman analizinde düzeltilmiş GAT değerleri ile diğer tonometreler arasında uyum mevcuttu. SKK ile GİB değerleri arasında korelasyon saptanmadı. Sonuç: Çalışmamızda tüm tonometrelerle elde edilen GİB değerleri arasında anlamlı farklılık saptanmadı. İcare topikal anestezi gerektirmemesi, kullanımının kolay olması, ölçümün hızlı olması nedeniyle klinikte uygun bir tonometre yöntemi olarak kullanılabilir. [ABSTRACT FROM AUTHOR] |
Copyright of Journal of Glaucoma-Cataract / Glokom-Katarakt is the property of Glokom-Katarakt and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
Database: | Academic Search Complete |
FullText | Links: – Type: pdflink Text: Availability: 0 CustomLinks: – Url: https://resolver.ebsco.com/c/xy5jbn/result?sid=EBSCO:a9h&genre=article&issn=13059173&ISBN=&volume=8&issue=4&date=20131201&spage=225&pages=225-230&title=Journal of Glaucoma-Cataract / Glokom-Katarakt&atitle=Normal%20G%C3%B6zlerde%20Goldmann%20Applanasyon%20Tonometre%20ile%20Rebound%20Tonometri%2C%20Dinamik%20Kontur%20Tonometre%2C%20Tonopen%20XL%27%C4%B1n%20Kar%C5%9F%C4%B1la%C5%9Ft%C4%B1r%C4%B1lmas%C4%B1.&aulast=%C3%96ZCAN%2C%20Altan%20Atakan&id=DOI: Name: Full Text Finder (for New FTF UI) (s8985755) Category: fullText Text: Find It @ SCU Libraries MouseOverText: Find It @ SCU Libraries |
---|---|
Header | DbId: a9h DbLabel: Academic Search Complete An: 93671198 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
IllustrationInfo | |
Items | – Name: Title Label: Title Group: Ti Data: Normal Gözlerde Goldmann Applanasyon Tonometre ile Rebound Tonometri, Dinamik Kontur Tonometre, Tonopen XL'ın Karşılaştırılması. – Name: TitleAlt Label: Alternate Title Group: TiAlt Data: Comparison of Rebound Tonometer, Dynamic Contour Tonometer, Tonopen XL with Goldmann Applanation Tonometer in Normal Eyes. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22ÖZCAN%2C+Altan+Atakan%22">ÖZCAN, Altan Atakan</searchLink><relatesTo>1</relatesTo><i> altanoz@gmail.com</i><br /><searchLink fieldCode="AR" term="%22ÇİLOĞLU%2C+Emine%22">ÇİLOĞLU, Emine</searchLink><relatesTo>2</relatesTo><i> drciloglu@gmail.com</i><br /><searchLink fieldCode="AR" term="%22ESEN%2C+Ebru%22">ESEN, Ebru</searchLink><relatesTo>3</relatesTo><i> ebrublg@yahoo.com</i><br /><searchLink fieldCode="AR" term="%22ÜNDAR%2C+İsmail+Hakkı%22">ÜNDAR, İsmail Hakkı</searchLink><relatesTo>4</relatesTo><i> dr_ismailundar@hotmail.com</i><br /><searchLink fieldCode="AR" term="%22HARBİYELİ%2C+İnan%22">HARBİYELİ, İnan</searchLink><relatesTo>5</relatesTo><i> inanharbiyeli@hotmail.com</i><br /><searchLink fieldCode="AR" term="%22ŞİMŞEK%2C+Firas%22">ŞİMŞEK, Firas</searchLink><relatesTo>6</relatesTo><i> firassimsek@gmail.com</i> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Journal+of+Glaucoma-Cataract+%2F+Glokom-Katarakt%22">Journal of Glaucoma-Cataract / Glokom-Katarakt</searchLink>. Dec2013, Vol. 8 Issue 4, p225-230. 6p. 3 Charts. – Name: Subject Label: Subject Terms Group: Su Data: *<searchLink fieldCode="DE" term="%22MEDICAL+equipment%22">MEDICAL equipment</searchLink><br />*<searchLink fieldCode="DE" term="%22TONOMETERS%22">TONOMETERS</searchLink><br />*<searchLink fieldCode="DE" term="%22COMPARATIVE+studies%22">COMPARATIVE studies</searchLink><br />*<searchLink fieldCode="DE" term="%22THICKNESS+measurement%22">THICKNESS measurement</searchLink><br />*<searchLink fieldCode="DE" term="%22ULTRASONICS%22">ULTRASONICS</searchLink><br />*<searchLink fieldCode="DE" term="%22STANDARD+deviations%22">STANDARD deviations</searchLink> – Name: Abstract Label: Abstract (English) Group: Ab Data: Purpose: This clinical study was conducted to compare intraocular pressure (IOP) measurements by Goldmann applanation tonometry (GAT), Rebound tonometry (Icare), Dynamic Contour tonometer (Pascal) and TonoPen XL (TP), to evaluate the effect of central corneal thickness (CCK) on IOP measurements. Materials and Methods: The study included 39 healthy subjects. IOP measurements were taken using Icare, Pascal, TP and GAT, respectively. CCK was measured with ultrasound pachymetry. To compare the measurements; the analysis of percentage change and Pearson correlation, the average, median, standart deviation and minimum-maximum values were calculated. The agreement between 2 different devices was assessed by use of the Bland-Altman method. Results: The average CCK was 528±16 μm (483-555 μm). The mean IOP, with Icare was 13.8±2.1 mmHg, with Pascal DCT was 14.0±2.2 mmHg, with TP was 14.0±3.4 mmHg, with GAT was 13.8±2.4 mmHg. There was no significant difference statistically between the IOP measurements of all methods (p>0.05). The percentage of the Icare, Pascal DCT and TP readings, which were all within ±3 mmHg of the GAT readings were respectively 66.6%, 61.5%, 56.4%. There was an agreement between corrected GAT values and the other tonometry values on the Bland-Altman graphs. There was no correlation between IOP measurements and CCT. Conclusion: In our study, there was no significant difference between the four methods of IOP measurements. Icare tonometry does not require topical anesthesia, is easy to use, the measurement is fast. So, it can be used as a relevant tonometry in clinical setting. [ABSTRACT FROM AUTHOR] – Name: Abstract Label: Abstract (Turkish) Group: Ab Data: Amaç: Bu klinik çalışmada Goldmann applanasyon tonometresi (GAT), Rebound tonometre (İcare), Dinamik kontur tonometre (Pascal DKT) ve TonoPen XL (TP) ile yapılan göz içi basınç ölçümlerinin (GİB) karşılaştırılması ve İcare, Pascal DKT, TP ile yapılan göz içi basınç ölçümleri üzerine santral kornea kalınlığının etkisini değerlendirmek amacıyla planlanmıştır. Gereç ve Yöntem: Otuz dokuz sağlıklı birey çalışmaya dahil edildi. Olguların sırasıyla İcare, Pascal DKT, TP ve GAT ile GİB ölçümleri yapıldı. Ultrasonik pakimetri kullanılarak santral kornea kalınlığı (SKK) değerleri elde edildi. Ölçümlerin karşılaştırılması için yüzde değişim analizi, Pearson korelasyon analizi ile ortalama, standart sapma değerleri ve minimum-maksimum değerleri hesaplandı. İki farklı metod arasındaki uyumluluk Bland-Altman metodu kullanılarak değerlendirildi. Bulgular: Çalışmaya dahil edilen olguların ortalama yaşı 38±4 (30-45) idi. Ortalama SKK 528±16 μm (483-555 μm). Ortalama GİB; İcare ile ortalama 13.8±2.1 mmHg, Pascal DKT ile ortalama 14.0±2.2 mmHg,TP ile 14.0±3.4 mmHg, GAT ile ortalama 13.8±2.4 mmHg, olarak ölçüldü. Ölçümler arasında anlamlı farklılık görülmedi. İcare, Pascal DKT, TP ile yapılan ölçümlerin yüzde kaçının GAT ile elde edilen değerlerin ±3 mmHg aralığında olduğu araştırıldığında sırasıyla %66.6, %61.5, %56.4 değerler elde edildi. Bland-Altman analizinde düzeltilmiş GAT değerleri ile diğer tonometreler arasında uyum mevcuttu. SKK ile GİB değerleri arasında korelasyon saptanmadı. Sonuç: Çalışmamızda tüm tonometrelerle elde edilen GİB değerleri arasında anlamlı farklılık saptanmadı. İcare topikal anestezi gerektirmemesi, kullanımının kolay olması, ölçümün hızlı olması nedeniyle klinikte uygun bir tonometre yöntemi olarak kullanılabilir. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Journal of Glaucoma-Cataract / Glokom-Katarakt is the property of Glokom-Katarakt and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
PLink | https://login.libproxy.scu.edu/login?url=https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&scope=site&db=a9h&AN=93671198 |
RecordInfo | BibRecord: BibEntity: Languages: – Code: tur Text: Turkish PhysicalDescription: Pagination: PageCount: 6 StartPage: 225 Subjects: – SubjectFull: MEDICAL equipment Type: general – SubjectFull: TONOMETERS Type: general – SubjectFull: COMPARATIVE studies Type: general – SubjectFull: THICKNESS measurement Type: general – SubjectFull: ULTRASONICS Type: general – SubjectFull: STANDARD deviations Type: general Titles: – TitleFull: Normal Gözlerde Goldmann Applanasyon Tonometre ile Rebound Tonometri, Dinamik Kontur Tonometre, Tonopen XL'ın Karşılaştırılması. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: ÖZCAN, Altan Atakan – PersonEntity: Name: NameFull: ÇİLOĞLU, Emine – PersonEntity: Name: NameFull: ESEN, Ebru – PersonEntity: Name: NameFull: ÜNDAR, İsmail Hakkı – PersonEntity: Name: NameFull: HARBİYELİ, İnan – PersonEntity: Name: NameFull: ŞİMŞEK, Firas IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 12 Text: Dec2013 Type: published Y: 2013 Identifiers: – Type: issn-print Value: 13059173 Numbering: – Type: volume Value: 8 – Type: issue Value: 4 Titles: – TitleFull: Journal of Glaucoma-Cataract / Glokom-Katarakt Type: main |
ResultId | 1 |