Factors influencing the size distribution of aerosols from jet nebulisers.

Bibliographic Details
Title: Factors influencing the size distribution of aerosols from jet nebulisers.
Authors: Clay, M M, Pavia, D, Newman, S P, Clarke, S W
Source: Thorax; Oct1983, Vol. 38 Issue 10, p755-759, 5p, 1 Chart, 3 Graphs
Abstract: The size distribution of saline and bronchodilator (terbutaline) aerosol droplets generated from four widely used jet nebulisers (Acorn, Upmist, Turret, and Inspiron Mini-neb) has been measured with a Malvern 2200 Laser Particle Sizer. The mass median diameter of aerosol droplets generated by each nebuliser was strongly influenced by the driving flow rate of compressed air. By increasing the flow rate from 4 to 8 1 min-1 mass median diameters were halved (p less than 0.01) and there was an increase in the mass of aerosol within the optimum respirable range (less than 5 micron). To achieve this range the following individual flow rates were required: Turret 4 1 min-1, Acorn and Upmist 6 1 min-1, and Inspiron Mini-neb 8 1 min-1. A significant inverse relation (p less than 0.001) was found between mass median diameter and the geometric standard deviation, indicating that the aerosols were smaller but more heterodisperse at high flow rates. Changes in drug concentration had little effect on aerosol size. In 72% of the nebulisations followed to dryness there was no significant change in mass median diameter during the course of nebulisation and in the remainder it was less than 1.3 micron. [ABSTRACT FROM AUTHOR]
Copyright of Thorax is the property of BMJ Publishing Group and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Complementary Index
FullText Text:
  Availability: 0
CustomLinks:
  – Url: https://resolver.ebsco.com/c/xy5jbn/result?sid=EBSCO:edb&genre=article&issn=00406376&ISBN=&volume=38&issue=10&date=19831001&spage=755&pages=755-759&title=Thorax&atitle=Factors%20influencing%20the%20size%20distribution%20of%20aerosols%20from%20jet%20nebulisers.&aulast=Clay%2C%20M%20M&id=DOI:
    Name: Full Text Finder (for New FTF UI) (s8985755)
    Category: fullText
    Text: Find It @ SCU Libraries
    MouseOverText: Find It @ SCU Libraries
Header DbId: edb
DbLabel: Complementary Index
An: 66047453
RelevancyScore: 819
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 818.93701171875
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: Factors influencing the size distribution of aerosols from jet nebulisers.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Clay%2C+M+M%22">Clay, M M</searchLink><br /><searchLink fieldCode="AR" term="%22Pavia%2C+D%22">Pavia, D</searchLink><br /><searchLink fieldCode="AR" term="%22Newman%2C+S+P%22">Newman, S P</searchLink><br /><searchLink fieldCode="AR" term="%22Clarke%2C+S+W%22">Clarke, S W</searchLink>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: Thorax; Oct1983, Vol. 38 Issue 10, p755-759, 5p, 1 Chart, 3 Graphs
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: The size distribution of saline and bronchodilator (terbutaline) aerosol droplets generated from four widely used jet nebulisers (Acorn, Upmist, Turret, and Inspiron Mini-neb) has been measured with a Malvern 2200 Laser Particle Sizer. The mass median diameter of aerosol droplets generated by each nebuliser was strongly influenced by the driving flow rate of compressed air. By increasing the flow rate from 4 to 8 1 min-1 mass median diameters were halved (p less than 0.01) and there was an increase in the mass of aerosol within the optimum respirable range (less than 5 micron). To achieve this range the following individual flow rates were required: Turret 4 1 min-1, Acorn and Upmist 6 1 min-1, and Inspiron Mini-neb 8 1 min-1. A significant inverse relation (p less than 0.001) was found between mass median diameter and the geometric standard deviation, indicating that the aerosols were smaller but more heterodisperse at high flow rates. Changes in drug concentration had little effect on aerosol size. In 72% of the nebulisations followed to dryness there was no significant change in mass median diameter during the course of nebulisation and in the remainder it was less than 1.3 micron. [ABSTRACT FROM AUTHOR]
– Name: Abstract
  Label:
  Group: Ab
  Data: <i>Copyright of Thorax is the property of BMJ Publishing Group and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://login.libproxy.scu.edu/login?url=https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&scope=site&db=edb&AN=66047453
RecordInfo BibRecord:
  BibEntity:
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 5
        StartPage: 755
    Titles:
      – TitleFull: Factors influencing the size distribution of aerosols from jet nebulisers.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Clay, M M
      – PersonEntity:
          Name:
            NameFull: Pavia, D
      – PersonEntity:
          Name:
            NameFull: Newman, S P
      – PersonEntity:
          Name:
            NameFull: Clarke, S W
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 10
              Text: Oct1983
              Type: published
              Y: 1983
          Identifiers:
            – Type: issn-print
              Value: 00406376
          Numbering:
            – Type: volume
              Value: 38
            – Type: issue
              Value: 10
          Titles:
            – TitleFull: Thorax
              Type: main
ResultId 1